Model-specific Specifications
Weighing Data Signum Advanced SIWADCP-1,-2,-3
Non-verifiable models
Weighing capacity (kg) 7 16 35 65
Calibration weight (g) 5,000 5,000 10,000 20,000
Accuracy class F2 F2 F2 F2
Resolutions for Signum Advanced
Non-verifiable models
-S Resolution >60.000d
-H Resolution >100.000d
-I Resolution >30.000d
Weighing Data Signum Surpreme SIWSBBP-1,-2,-3
Non-verifiable models
Weighing capacity (kg) 0.620 0.6/3.1 0.6/3.1 3.1 1.2/6.1 1.2/6.1 6.1 1.2/6.1 1.2/6.1
Readability (g) 0.001 0.01/0.1 0.01/0.1 0.01 0.01/0.1 0.01/0.1 0.01 0.01/1 0.1/1
Resolution code -H -D -P -H -D -P -H -K -T
Calibration weight (g) 500 2,000 2,000 2,000 5,000 5,000 5,000 500/5,000 5,000
Accuracy class E2 E2 E2 E1 E2 E2 E2 F1 F1
Verifiable/verified models
Model BD SI 200 BD SI 200 BD SI 200 BD SI 200 BD SI 200 BD SI 200 BD SI 200 BD SI 200 BD SI 200
EC type approval no. D07-09-010
Weighing capacity (kg) 0.620 0.600/3.1 0.600/3.1 3.1 1.2/6.1 1.2/6 .1 6.1 1.2/6.1 1.2/6.1
Readability (g) 0.001 0.01/0.1 0.01/0.1 0.01 0.01/0.1 0.01/0.1 0.01 0.1/1 0.1/1
Verification scale
interval e (g) 0.01 0.1 0.1 0.1 0.1 0.1 0.1 1 1
Resolution code -HCE -DCE -PCE -HCE -DCE -PCE -HCE -KCE -TCE
Calibration weight (g) 500 2,000 2,000 2,000 5,000 5,000 5,000 500/5,000 5,000
Genauigkeitsklasse E2 E2 E2 E1 E2 E2 F1 F1 F1
For all Models
Preload (kg) 033300000
Repeatability (g) 0.001 0.01 0.01 0.01 0.01 0.01 0.01 0.1 0.1
Linearity (g) 0.002 0.02 0.02 0.02 0.02 0.02 0.02 0.2 0.2
Ambient temperature
(for use in legal metrology only) 10°C….+40°C
Resolution for Signum Advanced
Non-verifiable Verified ex-works
-S Resolution >= 60,000 d -SCE Single range, Class II < 10,000 e, e = 10 d
-H Resolution > 100,000 d -HCE Single range, Class II > 10,000 e, e = 10 d
-K Resolution <= 50,000 d -KCE Dual range (fine range only for spec. capacity level)
(fine range only for spec. capacity level) Class II <= 5,000 e
-T Resolution <= 50,000 d -TCE Dual range (fine range can be continuously used)
(fine range can be continuously used) Class II <= 5,000 e
-D Resolution > 50,000 d -DCE Dual range (fine range only for spec. capacity level)
(fine range only for spec. capacity level) Class II > 5,000 e
-P Resolution > 50,000 d -PCE Dual range (fine range can be continuously used)
(fine range can be continuously used) Class II > 5,000 e
-I Resolution > 30,000 d -ICE Single range, Class II 30,000 e, e = d
97