Operation
47
  SQmin Function
 Purpose
  To display the allowable minimum 
sample quantity “SQmin” (sample 
quantity minimum) in accordance 
with the United States Pharmacopoeia 
(USP). According to USP guidelines, the 
uncertainty of measurement may not 
exceed 0.1 % of the sample quantity 
when substances are weighed with the 
highest degree of accuracy for volume 
determination. This additional function 
ensures that weighing results lie within 
defined tolerance limits corresponding 
to the requirements of your quality 
assurance system.
 Requirements
  The scale must be set up by a service 
technician to be able to use the 
SQmin function. The technician will 
determine the permitted minimum 
sample quantity and load this to your 
scale using the guidelines of your 
QA system. These settings cannot be 
changed by the user. He will document 
this setting via a “Weighing module 
test as per USP" certificate in which 
the measurements and min. sample 
quantity are logged. The SQmin 
function ensures that the weighing 
results correspond to USP guidelines.
 Characteristics
–  Displaying the minimum sample 
quantity: 
The value is shown in the text line for 
4 seconds after the k key is pressed. 
–  If the minimum sample quantity has 
not been reached: 
Symbol displayed: The ! symbol is 
displayed and weight values are marked 
with a “
!" in the printout.
–  GLP header: The minimum sample 
quantity entered for SQmin can be 
included on the printout.
  Parameter Factory Settings
  Display:      SQmin      Off
  Print in GLP header: Off
  SQmin  Display  *  No
        Yes
    Print in   *  No
    GLP header    Yes
  * = Factory setting
  See also the chapter on “Configuration: Application Parameters (Overview)"
§ Save the settings with the ) key and exit Setup: Press the ( key several times.