Getting Started
  Unpacking the Equipment
§ After unpacking the equipment, please 
check it immediately for any external 
damage.
$ If you detect any damage, proceed as 
directed in the chapter entitled “Care 
and Maintenance" under “Safety 
inspection."
$ Save the box and all parts of the 
packaging for any future transport. 
Unplug all connected cables before 
packing the equipment.
  Equipment Supplied
–  Complete scale
–  Operating instructions
–  Special accessories as listed on the bill 
of delivery, if ordered
 Setup
  Choose a location that is not subject to 
the following negative influences:
–  Heat (heater or direct sunlight)
–  Drafts from open windows and doors
–  Extreme vibrations during weighing
–  Excessive moisture
 Acclimatization
  Moisture in the air can condense 
on cold surfaces whenever the 
equipment is moved to a substantially 
warmer place. To avoid the effects of 
condensation, condition the scale for 
about two hours at room temperature, 
leaving it unplugged from the power 
supply.
  Equipment Downtime
  Switch off the equipment when not 
in use.
   Checking the Geographical Data 
Entered for Use in Legal Metrology 
(SIWR Models Only): 
  Preparation  
(See also the “Device Information” 
menu items listed under “Operating 
Menu Overview" in the chapter entitled 
“Configuration")
§ Press the e key to turn on the scale
§ While all segments are lit, press the 
T key
> 
Appl is displayed
§ To select “Device-specific information,"  
press k repeatedly; press T to 
confirm
§ To switch the display between 
information on weighing platform 1 
and weighing platform 2, press k 
repeatedly; press T to confirm 
>  View geographical data (configured 
prior to verification), for example:  
Latitude (in degrees): 
51 
4
  Altitude (in meters): 513 
5
  or 
  Gravitational acceleration (in m/s
2
): 
9.810 
6
  The scale can be used in legal 
metrology anywhere in Germany if the 
geographical data is as follows:
–  Latitude:  51.00 degrees
–  Altitude:  513 m
  This data corresponds to the following 
value:
–  Grav. acc.: 9.810 m/s
2
 
  These values are calculated for Germany 
based on a mean value for the Earth’s 
acceleration. The greater the precision 
of the geographical data entered, the 
greater the precision achieved with the 
weighing instrument; the tolerance 
range, however, is restricted accordingly 
(see above). 
  The tolerances ranges, for example for 
a scale with 3000e, are as follows:
–  ± 100 km for the latitude, and
–  ± 200 m for the elevation above seal 
level
! If used outside the specified zone, 
the scale must be re-verified for use 
in legal metrology: Please contact an 
authorized service technician.
5
Under-Scale Weighing 
SIW_BBP
A port for a under-scale weighing 
hook is located on the bottom of 
the weighing platform.
§ Remove the cover plate from the 
scale base (<-).
§ Screw in the under-scale weighing 
hook accessory.
§ Hang the sample from the hook 
(e.g. using a bent wire). 
(Install a draft shield if necessary).
Under-Scale Weighing SIW_DCP
A port for a under-scale weighing 
hook is located on the bottom of 
the weighing platform.
§ Remove the cover plate from the 
scale base (<-).
§ Screw in the under-scale weighing 
hook accessory.
§ Hang the sample from the hook 
(e.g. using a bent wire). 
(Install a draft shield if necessary).
! Under-scale weighing cannot be used 
with the SIWR... and SIWABBP.
  Signum under-scale weighing hooks 
can be ordered using option U1.
<
<