Configuring Printouts
81
 
Configuring Printouts
 Purpose
  You can configure individual printout 
formats for each application. Using the 
total data record for Totalizing and 
Net-total Formulation applications, you 
can define which parameters are printed 
using the c key. 
  In the “Print parameters” Setup menu, 
single, component and total data 
records can be configured, which 
contain the available print items for the 
respective applications. This should be 
carried out after setting the applications 
since some data in the printout is 
application-dependent.
 Characteristics
–  Quantity and extent of printout lists:
  6 lists each with a max. length of 
30 print items
  - Single printout Printer 1 
  - Component printout Printer 1
  - Total data printout Printer 1
  - Single printout Printer 2 
  - Component printout Printer 2
  - Total data printout Printer 2
–  Single, component and total data 
records can be configured separately
–  Print single printout: p key
  Auto printout of application when 
Setup menu is activated:
  – Animal weighing (averaging) 
– Checkweighing
  – Classification
  Signum
®
 2 and 3 only:
–  Print component printout: 
Totalizing/Net-total formulation 
with the O key (Setup: Application 
3: Totalizing: Printout: Component 
printout)
  Signum
®
 2 and 3 only:
–  Print totalizing printout: 
For selected application Totalizing/Net 
total formulation with the c key
–  When switching to another application 
in Setup, only the application-
dependent printout lists are deleted. 
The other printout lists remain saved.
–  Print items can be deleted individually: 
Press and hold the ( key
–  Print items “Form Feed” for record 
footer: 
Move to the next label start for printer 
type: YDP01IS: “Label” and YDP04IS, 
setting “Label, manual form feed”
–  ISO/GLP/GMP-compliant printout: 
The Setup menu configuration under 
“ISO/GMP-compliant printout” is also 
active for configured printouts.
 Preparation
§ Switch on the scale: Press the e key
§ While all segments are lit, press the 
) key
§ Select Setup: Press the k key 
repeatedly until setup is displayed
§ Select Config. printout: Press the 
) key
§ Press the k key repeatedly until 
prtprot is displayed
§ Press the ) key
  prtprot 
(see page 35 for a detailed menu list)
    7
      7.4   Header and ID header input
      7.5   Quantity, interface
1
)
      7.6  Standard, interface
1
)
      7.7  Component, interface
1
)
      7.8  Result, interface
1
)
      7.9  Quantity, interface
2
)
      7.10  Standard, interface
2
)
      7.11  Component, interface
2
)
      7.12  Result, interface
2
)
      7.13  ISO/GMP
      7.14  Date without time
1
)
      7.15   Automatic printout 
after stability
      7.16  Flex print
      7.17  Decimal separator
    9
      9.1  Factory settings
 = Signum
®
 2 and 3 only
–  The rows of the protocol list can be 
called up and activated individually. 
Example: see under Configuration,  
menu item 7.6
–  The print selection set as active appears 
with the left selection bar on the 
display, e.g. gross, tare, net.
§ Extend printout, press the ) key, the 
selection bar now appears on the right 
of the display.
§ Select print items using the k key
§ Apply the desired print items by 
pressing the ) key
§ Press the ( key to change the print 
selection set as active.  
The selection bar appears on the left.
  The required print item is set as active 
and appears in the printout.
1
)  Only available if the time module is present.
–  Print items can be deleted individually 
from the active printout selection: 
Press and hold the ( key
§ Save the settings with the ) key and 
exit Setup: Press the ( key several 
times.
  Additional Functions
  Printing the “Selection”and “List” 
Settings
–  LIST: Print the currently selected list 
SELECT: Print currently selectable items
§ When selection bar is in LIST or 
SELECT:  
Press the p key
>  Printout (example)
  INDIV.PRT
    List
  ====================
  Net (N)
  Gross (G#)
  Tare
  Tare (T2/PT2)
  Piece count
  ====================
  etc.