Data Interface
85
  Pin Assignment Chart
  COM1 and UniCOM female connectors: 
Round socket with screw lock hardware 
for model  
SIWSDCS/SIWSBBS, stainless steel 
version.
 
Front view
COM1 pin assignments
Round socket with screw lock hardware
Pin A:  Control output: “lighter”
Pin B:  Data output (TxD)
Pin C:  Data input (RxD)
Pin D:  Data terminal ready (DTR)
Pin E:  Internal ground (GND)
Pin F:  +5 V
Pin G:  Control output: “heavier”
Pin H:  Clear to send (CTS)
Pin J:  Control output: “equal”
Pin K:  Universal switch
Pin L:  Control output “set”
Pin M:  +12 V 
  Standard PS/2 pin assignment for 
SIWSDCS, SIWABBP, SIWSBBP, 
SIWSBBS 
Round Socket
Pin 1: Keyboard_CLK
Pin 2: Keyboard_DATA
Pin 3: Not used
Pin 4: Internal ground (GND)
Pin 5: +5 V
Pin 6: Not used
  For SIWDCP* appl. level 2 and higher
  Pin Assignment Chart
  COM1 female connectors: 
25-pin D-Submini female connector 
DB25S for model: SIW*DCP*.. 
COM1 pin assignments
Pin   1:  Shield
Pin   2:  Data output (TxD)
Pin   3:  Data input (RxD)
Pin   4:  Internal ground (GND)
Pin   5:  Clear to send (CTS)
Pin   6:  Internally connected
Pin   7:  Internal ground (GND)
Pin   8:  Internal ground (GND)
Pin   9:  Not used
Pin 10:  Not used
Pin 11:  +12 V for printer
Pin 12:  RES_OUT\
Pin 13:   +5 V (on/off for barcode 
scanner)
Pin 14:  Internal ground (GND)
Pin 15:  Universal switch
Pin 16:  Control output: “lighter”
Pin 17:  Control output: “equal”
Pin 18:  Control output: “heavier”
Pin 19:  Control output “set”
Pin 20:  Data terminal ready (DTR)
Pin 21:  Ground power supply (GND)
Pin 22:  Not used
Pin 23:  Not used
Pin 24:   Power supply +15...25 V  
(peripherals)
Pin 25:  +5 V
Required interface connector
(Recommended): 
25-pin D-Submini (DB25)  
with shielded cable clamp assembly and 
shield plate (Amp type 826 985-1C) and 
fastening screws  
(Amp type 164868-1) 
  Connecting a second weighing 
platform:
  On Signum
®
 2 and 3 models, you can 
connect a second weighing platform.  
You can use both the COM1 and 
UniCOM port.* 
  COM1 is operated in the RS-232 mode. 
A second weighing platform can use 
the following operating modes:
–  SBI
–  XBPI-232 (factory setting)
–  ADC-232
  UniCOM can be operated in either 
RS-232 or RS-485 mode. A second 
weighing platform can use the 
following operating modes:
–  SBI (RS-232 mode)
–  XBPI-232 (RS-232 mode)
–  ADC-232 (RS-232 mode)
–  IS-485 (RS-485 mode, xBPI mode, 
factory setting)
–  ADC-485 (RS-485 mode)
–  2. ADU via additional interface  
(menu: comspec)
  Pin Assignment Chart SIWAEDG
  COM1 connection: 
  The COM1 interface is compatible with 
the standard Signum COM1 in function; 
however, it uses a screw terminal.
PIN Function
1 CTS
2 DTR
3 RxD
4 TxD
5 GND
6 UNIV_IN
7 MIN
8 PAR
9 MAJ
10 SET
11 LOAD_Printer
12 RESET_OUT
13 GND
14 GND
15 5V_OOUT
16 5V_SWITCH
17 KBD_DATA
18 KBD_CLK
19 SHLD
20 LINE_OUT
* Not possible for SIWAEDG