Checkweighing Toward Zero Application
 
68
Checkweighing Toward Zero Application 
Example: 2
Checkweighing samples with a target weight of 1250 g and a tolerance range from -10 g to +30 g (using the factory-set application parameters) 
Settings (different from the factory settings): 
Setup: Application 2: Checkweighing toward zero (parameter 4. 7. 2)
Setup: Printout, PRTPROT 7.6, then select the menu line items of your choice
O
  1.)  Begin input of target and 
tolerance values
  2.)  Place a sample with the target 
weight (in this example, 
1250 g) on the platform
O
  3.) Save target value
1240
  4.)  For Signum
®
 3 only: 
Enter value for lower limit 
(in this example, 1240 kg)
O
  5.)  Save value for the lower limit
1280
  6.)  For Signum
®
 3 only: 
Enter value for the upper limit 
(in this example, 1280 kg)
O
     Save value for upper limit
  7.) Weigh samples
p
  8.)  Print the results 
Note: If automatic printout of 
results is enabled, you do not 
need to press the p key.  
The results are printed 
automatically.
Setp  +   1.250 kg    Target
Min   +   1.240 kg    Minimum
Max   +   1.280 kg    Maximum
G#    +   1.256 kg    Gross Weight
T     +   0.000 kg    Tare weight
N     +   1.256 kg    Net Weight
Lim   +    0.48 %     Percentage of deviation from 
target*
W.Diff+   0.006 kg     Absolute deviation from target
-------------------
  *   When displayed in relation to 
target value:  
If the weight is lighter than 
the target, the display shows: 
LL
     If the weight is heavier than 
the target, the display shows: 
HH