91
 
GMP-compliant 
Printouts
  When the corresponding menu item is 
active, the printout is bracketed by a 
GMP header and a GMP footer (GMP: 
“Good Manufacturing Practice”).  
The GMP header precedes the first 
measured result. The GMP footer is 
printed after the last result in a series 
of measurements (“GMP-compliant 
printout always for several application 
results,” menu item 7.13.3).  
To end a series of measured results, 
press and hold the p key. In this case, 
the A symbol  
is displayed after the GMP header is 
printed and remains in the display until 
the GMP footer is printed.
  If you toggle to a different platform 
(Signum
®
 2 and 3 only) while a GMP 
printout of several measured results is 
being generated (menu item 7.13.3), 
the GMP footer for the platform used 
up to that point is generated when you 
press the n key. The GMP header for 
the other platform is included on the 
next printout generated.
  A GMP-compliant printout is generated 
automatically at the conclusion of 
calibration/adjustment, linearization 
routines, as well as when you set or 
clear a preload.
Three examples of GMP headers and one example of a footer are shown in the following.  
On Signum
®
 1 models, the “date and time” line is not included.
Weighing platform WP 1:
-------------------  Dash line
14.01.2011    09:43  Date/Time 
1
) 
Type           SIWR  Signum
®
 Type
Ser.no.    12345678  Signum
®
 Serial no.
Vers.   1.1007.12.1  Software version of application
BVers.     01-25-01  Software version of basic version
-------------------  Dash line
  Weighing platform WP 2 (xBPI protocol): 
2
) 
-------------------  Dash line
14.01.2011    09:45  Date/Time 
1
) 
Type           SIWR  Signum
®
 Type 
Ser.no.    12345678  Signum
®
 serial no.
Vers.   1.1007.12.1  Software version of application
BVers.     01-25-01  Software version of basic version
Type       IS12000S  Platform type
Ser.No     12345678  Platform serial no.
-------------------  Dash line
  Weighing platform WP 2 (SBI protocol): 
2
) 
-------------------  Dash line
14.01.2011    09:45  Date/Time 
1
) 
Type           SIWR  Signum
®
 Type
Ser.no.    12345678  Signum
®
 serial no.
Vers.   1.1007.12.1  Software version of application
BVers.     01-25-01  Software version of basic version
Type            SBI  (Platform type)
-------------------  Dash line
  GMP footer:
-------------------  Dash line
14.01.2011    09:45  Date/Time 
1
) 
Name:                 Field for signature
   Space line
-------------------  Dash line
  1
) Only if clock module is available
  2
) Only for Signum
®
 2 and 3 display and control units