EasyManua.ls Logo

Sartorius SIWRDCP - Example: Defining Three Classification Classes

Sartorius SIWRDCP
124 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Signum 1-2-3
Classification Application
W
71
Example:
Defining three classes.
Settings (different from the factory settings):
Setup: Application 2: Classification
Setup: Printout, PRTPROT 7.6, then select the menu line items of your choice
O
1.) Begin input of
class delimiters
110
2.) For Signum
®
3 only:
Enter the upper limit for
Class 1 using the keypad
(in this example, 110 g)
O
3.) Save the upper limit for
Class 1
130
4.) For Signum
®
3 only:
Enter the upper limit for
Class 2 using the keypad
(in this example, 130 g)
O
5.) Save the upper limit for
Class 2
6.) Place the sample on the
weighing platform
Read the result
p
7.) Print the results
Note: If automatic printout of
results is enabled, you do not
need to press the p key.
The results are printed
automatically.
Lim1 + 0.110 kg Configured printout:
Lim2 + 0.130 kg see page 81
G# + 0.118 kg
T + 0.000 kg
N + 0.118 g
Class 2
-------------------
+