26
SBS-6500 User Manual
Storage Battery Systems, LLC.
When the next measurement is taken while in auto hold, the auto hold function is
restarted.
2) Stop Auto Hold Function
To stop the auto hold function, press the key.
3) AVG Clear
In the auto hold function, the AVG (average) feature is automatically activated.
AVG measurement function calculates AVG value from a multiple of repeated
measurements (maximum of 100 measurements) of battery measurement deviations.
AVG measurement is available only for the main measurement of the unit “Ω” or “V”.
In impedance (internal resistance) mode, main measurement (main display) “Ω” is the
AVG measurement. Sub measurements (sub display) of voltage or temperature are not
averaged.
In voltage mode, voltage will be the AVG measurement. Other sub display
measurements cannot be averaged.
To stop or reset AVG measurement, press key. Then AVG value is
cleared and ready to begin averaging again.