56
4.2.1 AERONEB SPECIFICATIONS
FINE MIST
ANP-1100
STANDARD
MIST
ANP-1000
Particle Size (VMD – Volume Median Diameter) 3.5 μm 5 μm
Nominal Nebulization Output Rate > 0.1 mL/min > 0.2 mL/min
Residual Volume < 0.2 mL < 0.2 mL
Mass Median Aerodynamic Diameter (MMAD) ~ 1.8 μm 2.5-3.0 μm
Geometric Standard Deviation (GSD) 2.0 μm 2.0 μm
STANDARDIZING MULTIPLE NEBULIZERS
Adjusting the nebulizer settings can allow two or more similar nebulizers to operate at
approximately the same output rate. For example: If nebulizer A has a slightly higher output
than nebulizer B when run at default duty cycle (50%), you can adjust the settings to effectively
obtain similar outputs from both.
To standardize multiple nebulizers, use the following procedure:
1) Determine the actual aerosol output rate for each nebulizer by characterizing the
nebulizers in
flexiWare
.
2) Use the following to determine the new nebulization rate (r
Neb
) at which one nebulizer will be
operated to match the other one:
Aao
Bao
BNebANeb
r
r
rr

Table of Contents