5.3 FX ADAPTER AND Y-TUBING
The FX adapter and Y-tubing come into contact with aerosol particles, mucus and humidity.
Routine cleaning of both is necessary to prevent particle accumulation on the interior surfaces.
To clean the FX adapter and Y-tubing, proceed as follows:
Prepare a solution of mild liquid dish soap and water.
Detach the adapter from the module.
Submerge the adapter in the solution and agitate to ensure that the solution enters all
three flow pathways.
Soak if necessary.
Remove the adapter from the soapy solution and rinse thoroughly with clean water.
Use dry compressed air to thoroughly dry the three adapter flow pathways.
Be sure no droplets remain in the small, middle pathway to the module as this can
result in damage to the module when the adapter is subsequently used for an
experimentation session.
The Y-tubing that connects subjects to the FX adapter is a consumable item that must be
replaced periodically. The initial instrument purchase includes appropriate tubing for each
module that is ordered. Additional Y-tubing kits may be ordered when the initial supply is
depleted.
5.4 EXTERNAL SURFACES
To clean the external surfaces of the
flexiVent
FX, simply run a damp, non-abrasive cloth across
the surface, using a mild solution of water and liquid dish soap if necessary.
When cleaning the clear panel that covers the piston (Lexan material), use only a clean, non-
abrasive cloth. Do not use Windex, Pledge, Isopropyl Alcohol or any alcohol-based cleaning
products. Similarly, do not use harsh soaps.