SPECIFICATION CONDITIONS MIN TYP MAX UNITS NOTES
Used volume
Modules FX1 – FX5
1.5 51 mL 1
Ventilation
frequency
Based on 1:1 I:E (max
varies with subject
weight)
6 600 br/min 6
Ventilation tidal
volume
Modules FX1 - FX5 10 mL/kg 5
Subject size Modules FX1 - FX5 8 5,100 g 1
Control modes
Software controlled mechanical and pressure controlled
ventilation
Gas supply Room air or non-flammable mixed gas
I:E ratio 1:10 2:3 2:1 6
Number of
subjects
measured
simultaneously
1 4 2
PEEP 0 3 20 cmH
2
O 8
Pressure range -80 80 cmH
2
O 7
Respiratory Mechanics
Oscillation
frequency
Up to 20% full stroke DC 50 Hz
Oscillation
amplitude
3-20Hz 10 mL/kg
up to 50 Hz 1 mL/kg