Reference Sheet Version #003, 11-03-08; Firmware Version 2.1 and later
2
Command Instructions and List (see manual for complete list and descriptions)
Input commands in upper or lower case letters and register commands by pressing Enter key.
19plus V2 sends an error message if an invalid command is entered.
(if OutputExecutedTag=N) If 19plus V2 does not return S> prompt after executing a command, press Enter key to get S>.
If new command is not received within 2 minutes after completion of a command, 19plus V2 returns to quiescent (sleep) state.
If in quiescent (sleep) state, re-establish communications by clicking Connect in Communications menu or pressing Enter key.
Category Command Description
GetCD
Get and display Configuration Data.
GetSD
Get and display Status Data.
GetCC
Get and display Calibration Coefficients.
GetEC
Get and display Event Counters.
ResetEC
Reset Event Counters.
GetHD
Get and display Hardware Data.
DS
Display status and setup parameters.
Status
DCal
Display calibration coefficients.
DateTime=mmddyyyyhhmmss
Set real-time clock month, day, year, hour, minute, second
BaudRate=x
x= baud rate (600, 1200, 2400, 4800, 9600, 19200, 38400, 57600, 115200). Default 9600.
Echo=x
x=Y: Echo characters as you type. x=N: Do not.
OutputExecutedTag=x
x=Y: Display XML Executing and Executed tags. x=N: Do not.
BatteryType=x
x=alkaline: Alkaline. x=nicad: Nickel-Cadmium. x=nimh: Nickel Metal Hydride.
InitLogging
After uploading data, initialize logging to make entire memory available for recording.
SampleNumber=x
After uploading data, set SampleNumber=0 to make entire memory available for recording.
HeaderNumber=x
x= header and cast number for first cast when logging begins.
General
Setup
QS
Place in quiescent (sleep) state. Logging and memory retention not affected.
PType=x
x=1: Strain-gauge pressure. x=3: Quartz pressure.
Volt0=x Volt1=x Volt2=x
Volt3=x Volt4=x Volt5=x
x=Y: Enable external voltage (voltage 0, 1, 2, 3, 4, and 5).
x=N: Do not.
Pressure
& Voltage
Sensor
Setup
BioWiper=x
x=Y: Configuration includes ECO-FL fluorometer with Bio-Wiper. x=N: Does not.
SBE38=x
x=Y: Enable SBE 38 secondary temperature sensor. x=N: Do not.
GTD=x and DualGTD=x x=Y: Enable 1 or 2 GTDs (Pro-Oceanus Gas Tension Device). x=N: Do not.
TGTD
Measure Gas Tension Device(s), output 1 converted data sample for each GTD.
RS-232
Sensor
Setup
SendGTD=command
Send command to GTD (any command recognized by GTD) and receive response.
OutputFormat=x
x=0: Raw, Hex. x=1: Converted, Hex. x=2: Raw, decimal. x=3: Converted, decimal.
x=4: Pressure and scan number, Hex. x=5: Converted, decimal, XML.
OutputSal=x
x=Y: Output salinity (psu). x=N: Do not.
OutputSV=x
x=Y: Output sound velocity (m/sec). x=N: Do not.
Output
Format
OutputUCSD=x
x=Y: Output sigma-t (kg/m
3
), battery voltage, operating current (mA). x=N: Do not.
MP
Set to Profiling mode.
NAvg=x
x= number of samples to average. Default 1. Must be even number (2, 4, etc.) if 19plus V2 includes
optional Quartz pressure sensor.
MinCondFreq=x
x= minimum conductivity frequency (Hz) to enable pump turn-on.
PumpDelay=x
x= time (seconds) to wait after minimum conductivity frequency reached before turning pump on.
AutoRun=x
x=Y: Start / stop logging when external power applied / removed. x=N: Do not.
Profiling
Mode
Setup (no
effect if in
moored
mode)
IgnoreSwitch=x
x=Y: Ignore switch for starting/stopping logging. x=N: Do not.
MM
Set to Moored mode.
SampleInterval=x
x = interval (seconds) between samples (10 - 14,400).
NCycles=x
x= number of measurements to take and average every SampleInterval seconds. Must be even number
(2, 4, etc.) if 19plus V2 includes optional Quartz pressure sensor).
MooredPumpMode=x
x=0: No pump. x=1: Run pump for 0.5 seconds before each sample.
x=2: Run pump during each sample.
DelayBeforeSampling=x
x= time (seconds) to wait after switching on external voltage before sampling.
ParosIntegration=x
x= integration time (seconds) for optional Quartz pressure sensor (1 – 600; default 1).
Moored
Mode
Setup (no
effect if in
profiling
mode)
MooredTxRealTime=x
x=Y: Output real-time data. x=N: Do not.
StartNow
Start logging now.
StartDateTime=mmddyyyyhhmmss
Delayed logging start: month, day, year, hour, minute, second.
StartLater
Start logging at delayed start time.
Logging
Stop
Stop logging or waiting to start logging. Press Enter key before entering command.
GetSamples:b,e or DDb,e Upload data from scan b to scan e.
GetCast:x or DCx Profiling mode only. Upload data from cast x.
Data
Upload
GetHeaders:b,e or DHb,e Upload headers from header b to header e
SL
Output last sample from buffer and leave power on.
SLT
Output last sample from buffer, take new sample and store data in buffer. Leave power on.
TS
Take sample, store in buffer, output data. Leave power on.
TSS
Take sample, store in buffer and FLASH memory, output data, turn power off.
TSSOn
Take sample, store in buffer and FLASH memory, output data, leave power on.
Sampling
GetLastSamples:x
Output last x samples from FLASH memory.
124