Section 4: Deploying and Operating SBE 19plus V2
44
Status Commands (continued)
DCal Get and display calibration coefficients,
which are initially factory-set and should
agree with Calibration Certificates shipped
with 19plus V2.
Example: 19plus V2 with strain-gauge pressure sensor (user input in bold, command used to modify coefficient
in parentheses).
dcal
SeacatPlus V 2.1 SERIAL NO. 0003 20 Oct 2008 14:46:05 [DateTime=]
temperature: 20-jul-08 [TCalDate=]
TA0 = -3.178124e-06 [TA0=]
TA1 = 2.751603e-04 [TA1=]
TA2 = -2.215606e-06 [TA2=]
TA3 = 1.549719e-07 [TA3=]
TOFFSET = 0.000000e+00 [TOffset=]
conductivity: 20-jul-08 [CCalDate=]
G = -9.855242e-01 [CG=]
H = 1.458421e-01 [CH=]
I = -3.290801e-04 [CI=]
J = 4.784952e-05 [CJ=]
CPCOR = -9.570000e-08 [CPCor=]
CTCOR = 3.250000e-06 [CTCor=]
CSLOPE = 1.000000e+00 [CSlope=]
pressure S/N , range = 2000 psia: 14-jul-08 [PRange= (psia), PCalDate=]
PA0 = 0.000000e+00 [PA0=]
PA1 = 0.000000e+00 [PA1=]
PA2 = 0.000000e+00 [PA2=]
PTEMPA0 = 0.000000e+00 [PTempA0=]
PTEMPA1 = 0.000000e+00 [PTempA1=]
PTEMPA2 = 0.000000e+00 [PTempA2=]
PTCA0 = 0.000000e+00 [PTCA0=]
PTCA1 = 0.000000e+00 [PTCA1=]
PTCA2 = 0.000000e+00
[PTCA2=]
PTCB0 = 0.000000e+00 [PTCB0=]
PTCB1 = 0.000000e+00 [PTCB1=]
PTCB2 = 0.000000e+00 [PTCB2=]
POFFSET = 0.000000e+00 [POffset= (decibars)]
volt 0: offset = 0.000000e+00, slope = 1.000000e+00 [VOffset0=, VSlope0=]
volt 1: offset = 0.000000e+00, slope = 1.000000e+00 [Voffset1=, VSlope1=]
volt 2: offset = 0.000000e+00, slope = 1.000000e+00 [Voffset2=, VSlope2=]
volt 3: offset = 0.000000e+00, slope = 1.000000e+00 [Voffset3=, VSlope3=]
volt 4: offset = 0.000000e+00, slope = 1.000000e+00 [Voffset4=, VSlope4=]
volt 5: offset = 0.000000e+00, slope = 1.000000e+00 [Voffset5=, VSlope5=]
EXTFREQSF = 1.000000e+00 [ExtFreqSF=]