Section 4: Deploying and Operating SBE 19plus V2
39
Status Commands (continued)
GetCC Get and display calibration coefficients,
which are initially factory-set and should
agree with Calibration Certificates shipped
with 19plus V2.
Example: 19plus V2 with strain-gauge pressure sensor (user input in bold, command used to modify coefficient in parentheses)
getcc
<CalibrationCoefficients DeviceType = 'SBE19plus' SerialNumber = '01906003'>
<Calibration format = 'TEMP1' id = 'Main Temperature'>
<SerialNum>01906003</SerialNum>
<CalDate>19-Oct-08</CalDate> [TCalDate=]
<TA0>1.155787e-03</TA0> [TA0=]
<TA1>2.725208e-04</TA1> [TA1=]
<TA2>-7.526811e-07</TA2> [TA2=]
<TA3>1.716270e-07</TA3> [TA3=]
<TOFFSET>0.000000e+00</TOFFSET> [TOffset=]
</Calibration>
<Calibration format = 'WBCOND0' id = 'Main Conductivity'>
<SerialNum>01906003</SerialNum>
<CalDate>19-Oct-08</CalDate> [CCalDate=]
<G>-1.006192e+00</G> [CG=]
<H>1.310565e-01</H> [CH=]
<I>-2.437852e-04</I> [CI=]
<J>3.490353e-05</J> [CJ=]
<CPCOR>-9.570000e-08</CPCOR> [CPCor=]
<CTCOR>3.250000e-06</CTCOR> [CTCor=]
<CSLOPE>1.000000e+00</CSLOPE> [CSlope=]
</Calibration>
<Calibration format = 'STRAIN0' id = 'Main Pressure'>
<SerialNum>01906003</SerialNum>
<CalDate>27-Oct-08</CalDate> [PCalDate=]
<PA0>-5.137085e-02</PA0> [PA0=]
<PA1>1.550601e-03</PA1> [PA1=]
<PA2>7.210415e-12</PA2> [PA2=]
<PTCA0>5.154159e+05</PTCA0> [PTCA0=]
<PTCA1>2.560262e-01</PTCA1> [PTCA1=]
<PTCA2>-8.533080e-02</PTCA2> [PTCA2=]
<PTCB0>2.426612e+01</PTCB0> [PTCB0=]
<PTCB1>-7.750000e-04</PTCB1> [PTCB1=]
<PTCB2>0.000000e+00</PTCB2> [PTCB2=]
<PTEMPA0>-7.667877e+01</PTEMPA0> [PTempA0=]
<PTEMPA1>4.880376e+01</PTEMPA1> [PTempA1=]
<PTEMPA2>-4.555938e-01</PTEMPA2> [PTempA2=]
<POFFSET>0.000000e+00</POFFSET> [POffset= (decibars)]
<PRANGE>1.000000e+03</PRANGE> [PRange= (psia)]
</Calibration>
<Calibration format = 'VOLT0' id = 'Volt 0'>
<OFFSET>0.000000e+00</OFFSET> [VOffset0=]
<SLOPE>1.260977e+00</SLOPE> [VSlope0=]
</Calibration>
<Calibration format = 'VOLT0' id = 'Volt 1'>
<OFFSET>-4.728750e-02</OFFSET> [Voffset1=]
<SLOPE>1.259474e+00</SLOPE> [VSlope1=]
</Calibration>
<Calibration format = 'VOLT0' id = 'Volt 2'>
<OFFSET>-4.715313e-02</OFFSET> [Voffset2=]
<SLOPE>1.259946e+00</SLOPE> [VSlope2=]
</Calibration>
<Calibration format = 'VOLT0' id = 'Volt 3'>
<OFFSET>-4.772396e-02</OFFSET> [Voffset3=]
<SLOPE>1.260486e+00</SLOPE> [VSlope3=]
</Calibration>
<Calibration format = 'VOLT0' id = 'Volt 4'>
<OFFSET>-4.765417e-02</OFFSET> [Voffset4=]
<SLOPE>1.260014e+00</SLOPE> [VSlope4=]
</Calibration>
<Calibration format = 'VOLT0' id = 'Volt 5'>
<OFFSET>-4.744167e-02</OFFSET> [Voffset5=]
<SLOPE>1.260255e+00</SLOPE> [VSlope5=]
</Calibration>
<Calibration format = 'FREQ0' id = 'external frequency channel'>
<EXTFREQSF>1.000000e+00</EXTFREQSF> [ExtFreqSF=]
</Calibration>
</CalibrationCoefficients>
Notes:
DCal and GetCC responses contain
similar information, but in different
formats.
Dates shown are when calibrations
were performed.
46