Section 4: Deploying and Operating SBE 19plus V2
54
Logging Commands (continued)
StartNow Start logging now.
StartDateTime=
mmddyyyyhhmmss Set delayed logging start month, day, year,
hour, minute, and second.
StartLater Start logging at time set with
StartDateTime=. If you need to change
19plus V2 setup after StartLater has been
sent (but before logging has started), send
Stop, change setup as desired, and then
send StartLater again.
Stop Stop logging or stop waiting to start
logging (if StartLater was sent
but logging has not begun yet).
Press Enter key before sending Stop.
Example: Program 19plus V2 to start logging on 20 January 2008
12:05:00. (user input in bold)
STARTDATETIME=01202008120500
STARTLATER
Notes:
If using StartNow or StartLater to
start logging, the 19plus V2 must be
set to Moored mode (MM), or if in
Profiling mode (MP) must be set to
ignore the magnetic switch
(IgnoreSwitch=Y).
After receiving StartLater, the
19plus V2 displays waiting to
start at . . . in reply to
GetSD or DS. Once logging starts,
the DS reply displays logging.
If the delayed start date and time
has already passed when StartLater
is received, the 19plus V2 executes
StartNow.
If the delayed start date and time is
more than 90 days in the future
when StartLater is received, the
19plus V2 assumes that the user
made an error in setting the delayed
start date and time, and it executes
StartNow.
Notes:
You may need to send Stop several
times to get the 19plus V2 to
respond.
If in Profiling mode and
IgnoreSwitch=N, slide the magnetic
switch off or send Stop to stop
logging.
You must stop logging before
uploading data.
61