x (1000Ω/V)
Test Current Short Circuit <2mA
Display Range
0.05MΩ - 300MΩ
Measuring Range (EN 61557-2)
0.05MΩ – 300MΩ
Resolution
0.01MΩ maximum
Accuracy ±(5% + 5 digits)
Repeat tests as per IEC61557-2 Approx 4000
7.6 Rpe Voltage Measurement (4mm Terminals)
Display Range 30V – 440V
Voltage Measuring Range 30V – 440V DC
30V – 440V AC 50-60Hz
Resolution 1V
Accuracy ±(5% + 2 digits)
7.7 Operating Current (via AC/DC Current Clamp)
Display Range 0.1A – 40.0A
Current Measuring Range 0.1A – 40.0ADC
0.1A – 40.0AAC 50-60Hz
Resolution 0.1A
Accuracy ±(5% + 2 digits)
7.8 DC Power (via AC/DC Current Clamp)
Display Range 0.00kW – 40kW
Measuring Range 0.50kW – 40kW
Resolution 0.01kW
Accuracy ±(5% + 5 digits)
8 Environmental Conditions
The Solar PV200/210 has been designed to perform tests and measurements in a dry
environment.
Maximum barometric elevation for making measurements is 2000M.
Overvoltage category IEC 60664/IEC 61010, 300V Category III (Red and Black 4mm
terminals only).
Pollution degree 2 according to IEC 61010-1.
Protective system IP40 according to IEC 60529.
Electromagnetic compatibility (EMC). Interference immunity and emitted interference
conforming to IEC 61326-1.
Operating temperature range of 0°C to 40°C, without moisture condensation.
The Solar PV200/210 can be stored at any temperature in the range -25°C to +65°C
(relative humidity up to 90%). The batteries should be taken out of the instrument for
storage.