PV200/210 Operating Instructions
- 26 -
Appendix A
IEC61557-2: Insulation
Intrinsic error or
influence quantity
Reference conditions or
specified operating range
Designation code
Intrinsic error Reference conditions A
Position Reference position ±90
o
E
1
Supply voltage At the limits stated by the
manufacturer
E
2
Temperature 0
o
C and 40
o
C E
3
Operating Error
)EEE 1.15|A|(B
2
3
2
2
2
1
+++±=
IEC61557-4: Resistance of earth connection and equipotential bonding
Intrinsic error or
influence quantity
Reference conditions or
specified operating range
Designation code
Intrinsic error Reference conditions A
Position Reference position ±90
o
E
1
Supply voltage At the limits stated by the
manufacturer
E
2
Temperature 0
o
C and 40
o
C E
3
Operating Error
)EEE 1.15|A|(B
2
3
2
2
2
1
+++±=