EasyManua.ls Logo

Sensorex TX3000 - 4 Overview of TX3000 Ph;Orp Transmitter; Illustration of Terminal Function

Default Icon
37 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
PRODUCT INSTRUCTION MANUAL
Page 5 of 29
Part 4 TX3000 Overview (cont.)
GLASS
REF
GND
TP
TP
pH/
ORP
T
HI
LO
WASH
POWER
AC
INPUT
POWER
TX3000
only
2
4-20mA
4-20mA
1
+/-12V
+
+
+
-
-
-
-
4.2 Terminal Functions:
4.3 Description of terminal function:
1 & 2 - 100~240AC: Power supply terminal
3 & 4 - WASH: External wash relay terminal
5 & 6 - REL2: LO, External relay terminal low control
7 & 8 - REL1: HI, External relay terminal high control
9 - N/C
10 - 4~20mA – terminal /G: Temperature current output terminal -
11 - 4~20mA – terminal /G: Temperature current output terminal +
12 - 4~20mA - terminal: Master measurement current output terminal -
13 - 4~20mA - terminal: Master measurement current output terminal +
14 & 15 +/- 12V output
16 - T/P: temperature probe
17 - SG : Connect the other end of temperature probe, or used as ±12V ground potential. In two-wire distributing
system, there should be a short circuit between this terminal and REF (a short circuit jumper is attached from
the factory)
18 - REF: Coaxial shield connecting pH/ORP electrode signal wire
19 - NC: NC
20 - GLASS: Coaxial inner conductor= pH/ORP electrode signal wire
FIG 4-2
Form: INSTRTX3000-Rev A
Not
Used

Related product manuals