Part 8 Calibration
8.1 Calibration Security:
8.1.1 Code authorization: There is a two level password protection built in to the TX3000. The code authori-
zation of parameter setting is prior to the code authorization of calibration setting. Therefore, you can unlock
the calibration with your “parameter set-up password, or directly with your calibration set-up password.
8.1.2 Setting Code : See Section 7.2
8.2 Entering Calibration Mode:
In Measurement mode press "CAL" and "MODE" keys simultaneously. This will bring you to OS screen. You
can press "CAL" key at any time to return to Measurement mode.
Enter
Enter
Mode
+
Cal.
Mode
OR
NOTE:
Ct1: Calibration mode for TECH buers, 4.01, 7.00, 10.00
Cn1: Calibration Mode for NIST buers: 1.68, 4.01, 6.86, 9.18
Form: INSTRTX3000-Rev A
FIG 8-1