Part 8 Calibration (cont.)
8.5 Calibration of Pre-set NIST buers (Cn1): Use NIST buers and follow instructions in section 8.4 (same
as Ct1).
8.6 ORP Calibration: It is unnecessary to make regular calibration of ORP electrodes. it is only necessary to
use ORP Buer to check the electrode or adjust the deviation of electrical potential. Press "CAL" + "MODE"
keys simultaneously to adjust ORP zero-point electrical potential.
Form: INSTRTX3000-Rev A
Cal.
Mode
+
Clean the electrode completely, and put
it into the mV standard solution (B225 -
Zobell's Solution) Check the dierence
between the readout and buer solution.
Press "MODE" or "RIGHT ARROW" key to
set the digit until it is equal to the buer’s
standard mV value. Then, press "ENTER"
key to save it or press "CAL" key to return to
the measurement mode.
Mode
OR
Enter
Enter
FIG 8-4