Part 8 Calibration (cont.)
8.3 2- Point Calibration (Any Buer): Follow instructions in section 8.2 to enter Calibration mode until you
reach "CA1 Redy" screen.
Note: Single point cali-
bration only modies the
OS oset. The slope(SLP)
still stays as the last slope
calibration.
Enter
When measuring standard buer’s
mV value, the sign begins to
ash. After receiving a stable reading,
the display shows a similar pH value.
Press "MODE" or "RIGHT ARROW" key
to modify display value until it matches
the standard buer value.
Mode
OR
Enter
Enter
Enter
Press "CAL" key to exit for
completion of single point
calibration.
Cal.
Enter
Cal.
When measuring 2nd
standard buer’s mV value,
the sign begins to
ash. After receiving a
stable reading, the display
shows a similar pH value.
Press "MODE" or "RIGHT
ARROW" key to modify
display value until it
matches the standard
buer value.
Mode
OR