User Manual SERVOPRO Plasma Trace N
2
Analyzer
4.0 SPECIFICATIONS
9
Fully microprocessor-controlled, with 4 x 40 LCD display
Self-diagnostic system software
Auto-zero, auto-span calibration system (optional).
Provides maintenance-free system
Digital output:
4 digital outputs: 3 are for range indication and1 for system status monitoring
Relay contact rating: 30 Vrms, 42.4 V peak or 60 Vdc
Alarm output:
2 digital outputs for a specified value chosen by the user
Relay contact rating: 30 Vrms, 42.4 V peak or 60 Vdc
4-20 mA isolated output
Maximum load impedance: 600 at 20 mA DC
Isolation: 750 VDC or AC peak maximum
Serial port: RS-232 (optional)
Custom options: contact factory
CE and Y2000 compliant
CSA approval on demand
- The analyzer is rated in accordance to Over Voltage Category II, Pollution Degree 2
- Ingress protection: IP20.
The SERVOPRO PLASMA analyzer is only suitable to measure nitrogen in helium and argon.