User Manual SERVOPRO PLASMA Trace N
2
Analyzer
APPENDIX 2
69
WARNING
The cable connected to the mA output terminal shall not be longer than 30 m, nor shall it
leave the building in which the analyzer is installed without suitable isolation, nor shall it be
routed outside of the building in which the analyzer is installed.
WARNING
The rear terminal connector is protected against ESD by a protective plastic cover. When the
plastic cover is removed to connect a wire to the terminal, operator shall be properly
grounded to earth.
All relay output terminals are rated: 30 Vrms, 42.4 V peak or 60 Vdc.