Shimadzu AOC-20 4 Using the control module
Solvent flush - solvent + air + sample + (air)
Solvent flush w/o air - solvent + sample + (air)
Standard internal - solvent + air + standard sample + air + sample + (air)
Standard internal w/o air - solvent + standard sample + sample + (air)
Syringe injection speed
Injection speed of the syringe.
Select Solvent
Selects the appropriate solvent for solvent wash in case there are three
solvent vials on the sampler rack. It is necessary to enable this option in
the Shimadzu AOC- 20i/s Setup dialog using the Use three solvents
checkbox.
Air Suction
Check if you want autosampler to suck air into the needle after sample
suction. This can help prevent premature evaporation of part of the
sample, when the needle is warmed-up.