Shimadzu AOC-20 5 Sequence
5 Sequence
To use the autosampler in the Clarity sequence, following conditions
must be met:
The sequence must be saved
The sequence must be set as Active Sequence
All used methods must have the External Start enabled and the Autostop
time set
Method Setup - AS dialog in the used methods must be filled
The required values for starting vial (SV), ending vial (EV), number of
injections from vial (I/V) and injected volume are set in the sequence
table.
Standard sequence operation
After sequence start, the AS control method will be sent to the sampler
and the datastation will wait for the injection signal from sampler. After
receiving it, the run starts and after finishing, the method for next injection
will be sent to the sampler.
Note: Autosampler control is not available during Single Run operation.