4 Using the control module Clarity Control Module
4.2 Method Setup - AS
Fig 7: Method Setup - AS - Parameters
Dwell time
Waiting time after sample injection.
Valid range: 0 - 99.9 s
Rinse with sample
Sample washing times of the syringe before injection.
Valid range: 0 - 99
Rinse with solvent
Solvent washing times of the syringe after injection.
Valid range: 0 - 99
Viscosity time
Waiting time of the plunger after sample suction.
Valid range: 0 - 99.9 s
Pump syringe
Number of pumpings.
Valid range: 0 - 99
Plunger injection speed
Injection speed of the plunger.
Plunger suction speed
Plunger suction speed at sample suction.
Injection type
Injection type.
Standard - sample + (air)
- 8 -