121
1
3
5
2
4
6
7
8
12
14
13
9
10
11
77 ApplicationsApplications
Measuring with the Pass/fail evaluation Mode
1. Set the Pass/fail evaluation mode
^ [Pass/fail evaluation Mode Settings] (P.12 0 )
2. Take measurements in the Pass/fail evaluation mode
(1) Place the container on the pan, and then press [ 0/T].
(2) Weigh the sample.
Pass or fail will be judged in accordance with the following conditions.
Conditions Pass/fail
evaluation
Analog Bar and
Comparator Marks
Upper limit of
specified range
< Display Invalid
Upper limit of pass
range
< Display ≤
Upper limit of pass
range
HI
Lower limit of
judgment range
≤ Display ≤
Upper limit of judgment
range
OK
Lower limit of pass
range
≤ Display <
Lower limit of judgment
range
LO
Display <
Lower limit of pass
range
Invalid

Table of Contents