135
99
Connections and Communications with External EquipmentConnections and Communications with External Equipment
1
3
5
2
4
6
7
8
12
14
13
9
10
11
Comparator Function
Command Function Description
Targeted measurement mode
TRGT Targeted measurement setting
TARGET=XX.XXXX Target value setting Less that the weighing capacity
LIMIT=XX.XXXX Permissible pass range setting More than the minimum display
Pass/fail evaluation Mode
CHKW Pass/fail evaluation setting
OVR.RNG=XX.XXXX Maximum judgment limit setting Upper limit at which judgment is possible
Judgment not possible above this value. Less than the
weighing capacity.
HI.LIM=XX.XXXX Upper pass limit setting Upper Pass/Fail threshold
Lower pass limit < Upper pass limit < Upper judgment limit
LO.LIM=XX.XXXX Lower pass limit setting Lower Pass/Fail threshold
Lower judgment limit < Lower pass limit < Upper pass limit
UND.RNG=XX.XXXX minimum judgment limit setting Lower limit at which judgment is possible
Judgment not possible below this value.
Comparator Pass/Fail Output
GO Pass/Fail output (Above excessive
weight)
HL
Pass/Fail output (Excessive weight) HI
Pass/Fail output (Suitable weight) OK
Pass/Fail output (Insufficient) LO
Pass/Fail output (Below insufficient) LL
Sensitivity and Unit Registration
Command Function Description
Calibration
ICAL Sensitivity (Internal)
ECAL Sensitivity (External)
ECAL.W=XXX.XXXX Calibration standard weight value
setting (External)
XXX.XXXX: Set value (g)
Unit Registration
g g unit setting Switches between ON / OFF.
mg mg unit setting Switches between ON / OFF.
ct Carat unit setting Switches between ON / OFF.
mom
*1
Momme unit setting Switches between ON / OFF.
*1 Not applicable to a verified balance as a legal measurement instrument in the EU.
System Setting
Command Function Description
Balance Software
ID=XXXX Balance ID setting Default setting [0 0 0 0]
ID Balance ID loading XXXX: Alphanumerals (4 characters)
STATE Setting status output Outputs the status for all function settings
TIME Date/Time loading DATE _ 2015_ D ec.11
TIME_12.34.56
Loads the date and time from the built-in clock
User Management
LOGIN=XXXX: YYYY Logs in by specifying the user name
(XXXX) and password (YYYY)
Possible to log-in with a command
from the PC.
XXXX : User ID (maximum of 20 alphanumerals)
YYYY : Password 4 numeric
LOGOUT Log-out [STAND-BY] displayed upon log-in.
UID Acquires the ID for the user currently
logged in
IDXX
XX: 01 ~ 10 Log-IN ID
Displays USER 01 – 10 on the display menu.