3.2 System Administration Tool
3.2.5 External Modules
TOC-V Administrator’s Manual
28
3.2.5 External Modules
External Modules can launch external software modules linked to TOC-Control V to
run in the background before and after measurement. Use this function to import or
export data using LIMS, etc.
Select External Modules in the TOC-Control V Administrator window to display the
Background Processing dialog box.
This function can only be used by users having access rights to Background Processing
Settings.
Figure 3.20 Background Processing Dialog Box
Enable online ASCII Export
This function exports data in ASCII format. The data is exported after at least one
sample (one row in the sample table) has been measured. Data from TOC-Control V is
exported online in ASCII format.
Launch External Application
External applications can be initiated by the TOC-Control V software at 4 points in the
measurement sequence; before starting a sample run, before starting a new sample,
after finishing a sample and after the completing a sample run. The necessary external
applications include either external database initialization before each sample
measurement, or data output to a Laboratory Information Management System
(LIMS). Select the times for the external application to be launched by placing a check
mark in the corresponding check box. Enter the full path of the Executable File
including file name for the application to be launched, or click Search for File and
select the appropriate file.