2 Shimadzu User Authentication Tool
2-4
2) Maximum password age: (0 to 3650 days)
When the maximum age is expired since the latest password was set, the warning is
displayed.
If “0” is set here, the maximum age is indefinite, and the warning is not displayed.
3) Same password is permitted:
The previously set password is permitted.
If this item is set ineffective, the previous password is not available when the
password is changed.
If it is required to change the password periodically, set this item ineffective.
4) Passwords must meet complexity requirements:
A password must be consisted of both alphabet and the numeral characters.
5) User lockout duration: (0 to 9999 minutes)
If login has failed beyond the specified number of bad logon attempts by a user, the
user cannot log in for the specified duration. This function is offered to prevent
illegal login by the third party.
If “0” is set here, the user lockout function is ineffective, and any user can continue
the login operation endlessly without regard to the number of bad logon attempts.
6) User lockout after: bad logon attempts (0 to 999)
Set the number of bad logon attempts (password input error) permitted in one user.
If “0” is set here, the user lockout function is ineffective, and any user can continue
the login operation endlessly without regard to the number of bad logon attempts.
7) PC lockout duration: (0 to 9999 minutes)
If login has failed beyond the specified number of bad logon attempts by a personal
computer, the login operation from the personal computer is not permitted for the
specified duration. This function is offered to prevent illegal login attempted with
different user IDs by the third party.
If “0” is set here, the PC lockout function is ineffective, and the login operation can
be continued endlessly without regard to the number of bad logon attempts.
8) PC lockout after: bad logon attempts (0 to 999)
Set the number of bad logon attempts (password input error) permitted in one
computer.
If “0” is set here, the PC lockout function is ineffective, and the login operation can
be continued endlessly without regard to the number of bad logon attempts.
9) Description:
Enter a comment in up to 255 alphanumeric characters.
10) [Default] button
Click the [Default] button to adjust the ER/ER default values.
11) [Update] button
Click the [Update] button to update the changed set values.
12) [Exit] button