CLV620
Window material of the view‐
ing window
Plastic
Mounting elements 2 tapped blind holes M5
Alternative: 4 fixing holes, Ø 5.5 mm
Laserwarnschild Lasered on
Safety EN 62368-1: 2014-08
Enclosure rating IP69K (DIN 40050, Part 9)
Test conditions:
•
Water spray volume: 14 l/min ... 16 l/min
•
Water pressure/temperature: 10000 KPa (100 bar) / 80 °C
•
Flat jet nozzle distance: 100 mm ... 150 mm
•
Spray angle: 0°, 30°, 60°, 90°
•
Cycle: 30 seconds per position
•
Rotational speed of test specimen: 5 rpm
Protection class
(Class 3, EN 61140)
For operation in SELV systems (EN 60950-1) or ES1 systems (EN 62368-1)
Weight
2)
854 g, with double protective bushing
Dimensions (L x W x H)
1)
85 mm x 84 mm x 154 mm
Conformity CE.
For further conformities and certificates, see the product page on the Internet at:
www.sick.com/CLV62x
1)
With digital outputs loaded.
2)
Without connecting cables.
3)
See also see "Design", page 13.
Ambient data
CLV620
EMC test Radiated emission: According to EN 61000-6-3: 2007-01
Shock resistance: According to EN 61000-6-2: 2005-08
Vibration resistance EN 60068-2-6: 2008-02
Shock resistance EN 60068-2-27: 2009-05
Ambient operating tempera‐
ture
0° C ... +40 °C
Storage temperature –20° C ... +70 °C
Permissible relative humidity 0% ... 90%, non-condensing
Ambient light immunity 2,000 lx, on bar code
Bar code print contrast (PCS) ≥ 60 %
7.2 CLV63x bar code scanner in the IP69K protective housing
NOTE
The relevant online data sheet for your product including technical data and dimen‐
sions can be downloaded from the Internet: www.sick.com/CLV63x
Features
CLV630 CLV631 CLV632
Working range Long range Mid range Short range
TECHNICAL DATA 7
8021479/19ZB/2021-10-29 | SICK T E C H N I C A L I N F O R M A T I O N | CLV62x, CLV63x and CLV64x with IP69K Protective Housing
33
Subject to change without notice