EasyManua.ls Logo

SICK Lector64x Flex - Page 29

SICK Lector64x Flex
96 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
typically 10°... 20°
Figure 7: Mounting angle to use, depending on the application
1
Typical angle 10° ... 20°
The typical value is between 10° and 20°.
Depending on the application, an angle of between 0° (bright field light) and 45° (dark
field light) may be advisable.
5.5.3 Field of view diagrams
Lector64x Flex
0
0 200 400 600 800 1000 1200 1400 1600 1800 2000 2200
Field of view in mm²
235 x 160
470 x 320
545 x 370
625 x 425
780 x 530
700 x 475
155 x 105
310 x 210
390 x 265
80 x 55
V2D642R
4
3
2
f = 12mm
f = 16mm
f = 50mm
f = 75mm
f = 25mm
f = 35mm
1D code
Working distance/focus position in mm
1
Min. resolution in mm
2D code
0.2
0.3
0.1
0.2
0.4
0.6
0.4
0.3
0.1
0.5
0.5
0.7
0.6
0.7
0.8
0.9
1
Field of view in mm
2
2
Minimum resolution in mm
3
1D code
working distance (max.)
4
2D code
field of view (mm x mm)
5
Working distance/Focus position in mm
MOUNTING 5
8016185/19E9/2020-10-21 | SICK O P E R A T I N G I N S T R U C T I O N S | Lector64x/65x Flex, Lector65x Dynamic Focus
29
Subject to change without notice

Table of Contents