EasyManua.ls Logo

SICK MCS100E HW - Check the Flow Meter

SICK MCS100E HW
46 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Preventive Maintenance
18 MCS100E HW Maintenance Manual 8009624/2009-09 (V1.4) ©SICKMAIHAKGmbH
3.7
Check the flow meter
Check the zero point
1 View the "bar graph" display (F2 key).
2 Switch the FID (Option) off (mains switch at the rear of the operating unit).
MCS100E HW switches automatically to stand-by.
3 Wait until all measured values are shown as "0“.
4 Close the zero gas supply for MCS100E HW and FID (Oprion) at the compressed air sta-
tion.
5 In the "flow" bar graph diagram, the flow meter must indicate a value slightly higher
than "0" (0 .. 5 l/h).
Caution: The indicator reaction is very slow. Turn the potentiometer slowly.
If exactly "0" is shown (then negative values can also be set), turn the potentiometer
slowly counterclockwise (higher values) until a higher value than zero is set and
then turn slowly clockwise until "0" is displayed.
If a higher value than 5 l/h is displayed, turn the potentiometer slowly clockwise
(lower values) until "0" is displayed.
Fig. 9 Potentiometer of LPM43
6 Open the zero gas supply.
7 If necessary, reconnect the FID (
P. 39, § 3.10)
8 Optional: Switch the FID on.
Wait until FID switches to operation mode. (The green LED lights).
9 Switch the MCS100E HW to measuring mode.
Optional: Press the F4 function key and acknowledge the FID error message.
–Press the F3 function key.
Select "System Start" (confirm safety prompt with "Yes").
Exit the menu with "ESC".
If you also want to perform a leakage test: Disconnect the FID from the gas
path (
P. 39, § 3.10), you then do not have to switch the FID off and you avoid
the MCS100E HW switching to "Störung/Error".
Lower potentiometer
¯higher values
¬ lower values

Related product manuals