3-6 February 5, 2004 FFDM-S Service Manual
DirectRay PXCM Installation
Mount the DirectRay PXCM in the mammography equipment on four size M4
studs, in the position indicated in Figure 3-3.
NOTE: To ensure reliable communications, keep the fiber optic connections covered by protective
caps when not in use.
Figure 3-3.
DirectRay PXCM Stud Hole Locations
Figure 3-4 shows the PXCM connectors used for the cable that connects the
Detector to the DirectRay PXCM, and the fiber optic cable to the DirectRay
Interface card.
Figure 3-4.
DirectRay PXCM Connectors (for cable to Detector and fiber optic cables to
Interface card)
GS20_412.cdr
Connector for cable
to DirectRay Detector
Fiber optic connectors
to DirectRay Interface card
Connector for cable
to Generator
Connector for PXCM
power cord
PXCM power switch