EasyManua.ls Logo

Siemens MAMMOMAT NovationDR - Determining the Half Value Layer (HVL)

Siemens MAMMOMAT NovationDR
95 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Calibrating the AEC February 5, 2004 4-5
Determining the Half Value Layer (HVL)
Select the Determine HVL procedure option if you need to determine the HVL of
the X-ray Generator for a given target/filter combination.
This procedure requires taking four exposures.
The first exposure and the last exposure are “control” exposures with no
attenuating material in place.
The second exposure is taken with Aluminum in the path of the X-ray beam.
For the Molybdenum target, use a 0.3-mm thickness of Aluminum. For the
Tungsten target, use a 0.4-mm thickness of Aluminum.
The third exposure is taken with a thicker amount of Aluminum in the path of
the X-ray beam.
- For the Molybdenum target and Molybdenum filter, use a 0.4-mm
thickness of Aluminum.
- For the Molybdenum target and Rhodium filter, use 0.5-mm.
- For the Tungsten target and Rhodium filter, use a 0.6-mm thickness of
Aluminum.
The result of the HVL procedure is a fractional number representing the HVL in
millimeters of Aluminum.
NOTE: If you already know the HVL and want to proceed with determining the ESE value, go to
“Determining the Entrance Skin Exposure (ESE)” on page 4-8.
To determine the HVL value:
1. In the tool’s main window, select the Determine HVL procedure option.
In the main window, the settings change and instructional text is displayed:
The HVL procedure has been selected
The units have been changed to R/mR
The default AGD is now displayed in mrad instead of mGy
Instructions for Generator and dosimeter configuration are displayed in
the text window
An example of the main window with the Determine HVL procedure option
selected is shown below.

Table of Contents

Other manuals for Siemens MAMMOMAT NovationDR

Related product manuals