Calibrating the AEC February 5, 2004 4-13
Alternatively, you may see the following output in the terminal window where you
started the tool:
Installing goaltab to long-term Brick configuration...
(/usr/local/bin/xrayconfig goaltab save /tmp/goaltab)
"save" OK.
Installing goaltab to effective and short-term Brick configuration...
(/usr/local/bin/xrayconfig goaltab load /tmp/goaltab)
*** Error 256 running rssio_exec for load.
**** unknown error
.SendGoalToBrick:Goal Table goaltab NOT loaded on Brick.
***************************************
*** Error sending goaltab to brick.
***************************************
If you see the “save OK” message but an error instead of the “load OK” message, be
assured that the goaltab has been stored in the PXCM’s long-term storage and will
be auto-installed on reboot. Simply reboot the PXCM before attempting to take
AEC exposures. For information on rebooting the PXCM, refer to “Rebooting the
PXCM” on page 5-21.
Copying Goal Images
You can copy an existing goaltab file to the PXCM under another name.
To copy the goaltab file to the PXCM using a different name, open the File
menu and choose Copy. The Copy Popup dialog box displays, indicating the name
of the source goaltab file (this is the goaltab for Mo/Mo filter/target combination).
Enter the copy’s name in the Copy file goal tab to: field, then click OK. The
goaltab for Mo/Mo will be copied to a new file on the PXCM, using the file name you
entered.