MOBILETT Plus/ -E/ -HP SPR8-220.031.04 Page 14 of 30 Siemens AG
Rev. 04 09.03 CS PS 24 Medical Solutions
2 - 14 Installation
Connection of plugs to the main CPU board, D1 2
1. Remove the protection plate of the CPU board below to expose the main CPU D1
(see Fig. 18).
2. Disconnect pin plug K9 from board D1 (see Fig. 18).
3. Place the disconnected K9 cable in between the chassis frame and the D4 board
(see Fig. 15).
4. Mount a flat tie holder according to pos. 1 / Fig. 15.
5. Place a cable tie (2.5x100mm) around the disconnected K9 cable according to
pos. 2 / Fig. 15.
6. Place the large shrinking tube (Part No. 66 11 115, 50 mm length) around the
loose K9 pin plug (see pos. 3 / Fig. 15).
7. Connect pin plug K9 to K9b of cable 65 61 992 (see Fig. 15).
8. Shrink the shrinking tubing (see pos. 3 / Fig. 15) over the connected pin plugs K9
and K9b.
MOB00464
K9
K9b
D4
Fig. 15 Positioning of cable tie and shrinking tubing
2
3
remote control
system
1

Other manuals for Siemens MOBILETT Plus HP

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Siemens MOBILETT Plus HP and is the answer not in the manual?

Siemens MOBILETT Plus HP Specifications

General IconGeneral
BrandSiemens
ModelMOBILETT Plus HP
CategoryMedical Equipment
LanguageEnglish

Related product manuals