MOBILETT Plus / - E/ - M SPR8-215.898.01 Page 6 of 6 Siemens-Elema AB
Rev. 05 01.02 SPS-UD Solna, Sweden
1 - 6 General
• Measure the capacitor bank voltage on the bus bar (Fig. 3).
• If one of the capacitor series of the capacitor bank is still charged (U > 10 V), it can be
discharged via resistor R103 on D7.
• Connect the voltmeter to the measurement points parallel to the insulated cables which
will be used to discharge the capacitor bank (U < 10 V) (Fig. 4).