2 - 1
Siemens-Elema AB SPR8-215.898.01 Page 1 of 22 MOBILETT Plus / - E/ - M
Solna, Sweden Rev. 05 01.02 SPS-UD
Display messages 2
Dis-
played
messages
Mode of
operation
Cause Possible reason
Err 3* During
exposure
Interrupted expo-
sure because of over
voltage >142 kV
Only with "Err 3".
• Check the filament heating value
for the selected kV, see section
"P07 - Read-out filament values
from battery backed RAM".
Compare the value with the value
in the delivery protocol. If the
difference is large, make a new
adaptation, see section "P10 - Fast
adaptation".
Err 3" in combination with:
• Err 21" & "Err 22".
Check the measuring point
"FIL_VAL" on D1, see page 160**
and page 110**.
• CAL 13". Check the 60 V-supply
voltage on D4, see page 160**.
* See section "Monitoring the kV- and mA-controls"
** Wiring diagram X037E
Tab. 1