MOBILETT Plus / - E/ - M SPR8-215.898.01 Page 2 of 22 Siemens-Elema AB
Rev. 05 01.02 SPS-UD Solna, Sweden
2 - 2 Display messages
Err 4* During
exposure
Measured kV value
deviates more than
5 kV from selected
kV.
The deviation can depend on the
adaptation.
Measure the kV curve at
measuring point "HTPEAK", see
page 140**. Compare measured
curve with the reference curve, see
page 141**.
For a large deviation:
Check the filament value, see
section "P07 - Read-out filament
values from battery backed RAM",
and compare it with the values in
the delivery protocol.
If the difference is large a new
adaptation is needed, see section
"P10 - Fast adaptation"
See also section "Error diagnosis
HT and single tank".
Check the connection K9 between
D1 and H1 (single tank) and the
connection between D1 and D13
(K8), see page 140**
Dis-
played
messages
Mode of
operation
Cause Possible reason
* See section "Monitoring the kV- and mA-controls"
** Wiring diagram X037E
Tab. 1