Display messages 2 - 5
Siemens-Elema AB SPR8-215.898.01 Page 5 of 22 MOBILETT Plus / - E/ - M
Solna, Sweden Rev. 05 01.02 SPS-UD
CAL 7 During
exposure
The exposure time is
too long.
The maximum exposure time is 2 s.
The exposure is normally
interrupted earlier by "Err 5".
Check the capacitance in the
capacitor bank, refer to section
"P03 - Capacitor bank and
charging test".
Check the filament heating, refer to
section "P01 - Filament test".
Check the tube current in
measuring point "MAPOS" on D1,
see page 140 and 141.
Check the current to the single
tank, see section "Error diagnosis
HT and single tank".
Err 9 During
exposure
When measuring the
kV value, a large
high tension asym-
metry is detected.
Normally the kV value is symmetric
around ground. The acceptable devi-
ation is ± 19.5 kV. This corresponds
to HTD 2.5 V± 0.5 V.
Intermittent measurement cables
or defective single tank.
Check the measurement cable K9
between D1 and the single tank,
refer to page 140**.
Check the current to the single
tank, refer to the section "Error
diagnosis HT and single tank".
Measure the HTD signal on D1 and
compare it with notes and curves
on page 140 and 141**.
Dis-
played
messages
Mode of
operation
Cause Possible reason
* See section "Monitoring the kV- and mA-controls"
** Wiring diagram X037E
Tab. 1