MOBILETT Plus / - E/ - M SPR8-215.898.01 Page 6 of 22 Siemens-Elema AB
Rev. 05 01.02 SPS-UD Solna, Sweden
2 - 6 Display messages
Use 10 Generator ON.
The message
does not inter-
rupt a running
exposure.
The single tank tem-
perature measured
is > 55°C or
< 15°C.
The generator is overheated and
must cool down or the generator
was stored in a cold room and must
warm up prior to use.
Too high or too low environmental
temperature.
As soon as the single tank
temperature is
< 50°C , check the measurement
circuit for the temperature; refer to
page 140**, contact K9.
CAL 12 During
preparation
Stored filament
value is lost.
Check the voltage in the backup
battery, refer to page 110**. The
battery should have a voltage
> 2.8 V
Check that the battery jumper is
securely attached, see page 110**.
Make a new adaptation, see
section "P10 - Fast adaptation".
CAL 13 Generator ON The voltage supply
(60 V DC) to the fila-
ment circuit is
beyond its limit.
Check fuse U6 on D8, refer to page
160**.
Check the filament supply voltage
+60 V DC, refer to page 160**.
Dis-
played
messages
Mode of
operation
Cause Possible reason
* See section "Monitoring the kV- and mA-controls"
** Wiring diagram X037E
Tab. 1