EasyManua.ls Logo

Siemens MOBILETT Plus - Page 97

Siemens MOBILETT Plus
114 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Mechanical systems 10 - 19
Siemens-Elema AB SPR8-215.898.01 Page 19 of 28 MOBILETT Plus / - E/ - M
Solna, Sweden Rev. 05 01.02 SPS-UD
Remove the all arm covers and the column covers.
Relieve the tension in the spring pack by loosening the adjustment screw above the tilt
bar.
Disconnect all cables to the collimator.
Remove the collimator.
Disconnect all electrical connections to the single tank.
Remove the supporting hub in the rotation axis of the fork.
Remove all cable clamps and cable ties on the arm and the chassis.
Disconnect the single tank and the fork from the arm.
Disassemble the fork so the cables can be removed.
Disconnect the chassis connections and remove the cables (Fig. 22).
Fig. 22
Danger of injury!
The arm may move uncontrollably when the weight of the fork
and single tank is removed.
Ensure that the arm is properly secured to the unit foot with the
rope.
Demount
Remount
WARNING

Table of Contents

Other manuals for Siemens MOBILETT Plus

Related product manuals