12 Imaging system overview
MOBILETT XP Digital SPR8-230.840.30.03.02 Siemens AG
12.05 CS SD 24
Page 12 of 68
Medical Solutions
Use the initialization code when the normal code in unable to start for some reason or
when you have lost the Ethernet settings. Using the initialization code sets the Ether-
net-related settings to the default settings and allows you to perform the start-up operation
for the initialization code.
In this case, the connection can be made by setting the Ethernet settings in the control PC
to the 50G default (*2).
This default connection allows you to upgrade the firmware again and also to check and
set the Ethernet settings again in order to recover the system.
Fig. 6: Power box installation switch
Initialization switch: Switch 1 mounted on PCB-50 XRY
Default settings: IP Address=192.168.100.11 (factory setting)