64 Miscellaneous
MOBILETT XP Digital SPR8-230.840.30.03.02 Siemens AG
12.05 CS SD 24
Page 64 of 68
Medical Solutions
Generator module configuration 0
The MOBILETT XP Digital is delivered with preconfigured generator module cgbbig3.dll
for communication between the imaging system and generator. Do not change these val-
ues.
Procedure for checking the generator module:
Select SYSTEM/CONFIGURATION/ADMINISTRATOR SETUP/MODULE/GENERA-
TOR.
Enter the administrator password.
¹ See password list.
Check CXDI GEN value: cgbig3 (Fig.8/p.64).
Check the generator communication module: valid(Fig.8/p.64).
Click “OK”.
Fig. 8: Generator setup