14 Error messages of the imaging system
MOBILETT XP Digital SPR8-230.840.30.03.02 Siemens AG
12.05 CS SD 24
Page 14 of 68
Medical Solutions
3-
2Error messages of the imaging system
General 0
Service messages and messages from the operator's manual (customer messages) are
included.
The following table shows the alert code window.
Fig. 7: Alert window
Area
#
Severity Auto-
matic
display
Blinking
of "ERR"
Meaning
1 Fatal
error
Yes Yes The whole screen blinks red and the alert
window appears automatically. The system
cannot be recovered. A service person must
be called.
2 Alert Yes Yes A red ERR key blinks and the alert window
appears automatically. The exposure can-
not be continued without a user operation.
3 Error Yes No A red ERR key is displayed and the alert
window appears automatically. The ERR key
does not blink.The chances of the user being
able to recover the system is better than for
an alert.