22 Error messages of the imaging system
MOBILETT XP Digital SPR8-230.840.30.03.02 Siemens AG
12.05 CS SD 24
Page 22 of 68
Medical Solutions
Error 0
Tab. 5 Service Error Table
Code Reply
keys
Message
--- OK #1:Error
#2:File Open Error:filename
#3:File is not found.
[Cause]
The file may have been damaged.
[Operation after the reply key is pressed]
-1 OK #1:Error
#2:Error
#3:
[Cause]
Undefined error. This error indicates that it is not defined by the
CCR manager. The CCR manager controls various tasks. If an
error not defined in tasks occurs, error code "-1" is assigned to
the error and displayed. This error rarely appears because most
errors have been defined. Take the necessary action while
referring to the system log for the CCR. Local undefined errors
output by the OPU can be identified since they are displayed in
the message box of the OS.
[Operation after the reply key is pressed]
After you confirm it, press OK.
-2 OK #1:Error
#2:Insufficient tmp disk space. Output the saved data or erase
the backed up data.
#3:
[Cause] The remaining space in the temporary storage directory
dttmp is less than the specified MB.
[Operation after the reply key is pressed]
Press OK after confirmation.