16 Error messages of the imaging system
MOBILETT XP Digital SPR8-230.840.30.03.02 Siemens AG
12.05 CS SD 24
Page 16 of 68
Medical Solutions
Fatal error 0
Tab. 1 Service Fatal Error Table
Code Reply
keys
Message
-100 OK, #1:Fatal error
#2:Sensor unit: Self-check error
#3:Detected an error.#4:ER(n) (xxxx) xxxx: Error code
[Cause]
An undefined error has been found in the sensor A/D board.
[Operation after the reply key is pressed]
Nothing
-102 -14,
-15,
-95,
-98
-14
Error in starting the task in DLL (failed to begin thread.)
-15
Error in DLL resource (driver does not start.)
-95
Number of sensor units is incorrect (no sensor unit is con-
nected.)
-98Capture device cannot be found.
-103 Shut
down
#1:Fatal error
#2:READ timeout error
#3:Failed to transmit the data.
[Cause]
Image transferring from Ethernet has failed.
[Operation after the reply key is pressed]
For the reply cannot be done, the system must be shutdown.