40 Error messages of the imaging system
MOBILETT XP Digital SPR8-230.840.30.03.02 Siemens AG
12.05 CS SD 24
Page 40 of 68
Medical Solutions
Warning 0
Tab. 7 Service Warning Table
Code Reply
keys
Message
-150 OK #1:Warning
#2:Sensor unit:Self-check error
#3:Detected an error.
#4:ER(n) (xxxx)
#4:"Firm PLD err(n) (eff1 or eff2)
0xeff1: Failure has been detected in the received Hex format file
while the Firm/PLD/DP files are downloading.
0xeff2: Failure has been detected in the received Hex format file
while the Firm/PLD are downloading. n: Sensor index (1 to 4)
xxxx: Error code
[Cause]
An undefined error has been found in the sensor A/D board.
[Operation after the reply key is pressed]
Nothing.
-151 OK
0xeffe
0xeffd
0xeffc
#1:Warning
#2:Sensor unit:Flash ROM error
#3:Detected flash ROM error. Call a service person.
[Cause]
A warning on the flash ROM has been detected in the sensor
A/D board.
[Operation after the reply key is pressed]
When the OK key is pressed, the screen returns to the normal
display. Imaging can be performed by touching the exposure
mode key again. Although imaging can be performed, a service
person needs to be called in.